Table C-1Diagnostic Log Codes
Code
xxx = any
alphanumeric
characters
4f4b
80007xxx
80008xxx
80009xxx
8000axxx
113
The diagnostic codes given in C-1 may appear in an SC 7000 or SC
9000XL Diagnostic Log, and help in troubleshooting a malfunctioning
Monitor. Both possible cause and suggested remedial action are listed for
the field-significant codes.
Only codes for field-related procedures are given in the Table. There are
also a number of other diagnostic codes, not listed in the Table, that have
significance for engineering in helping to improve product performance. If
observed, report these codes to the factory. Refer to Chapter 1, Section
1.11, for the procedure to download the Diagnostic Log to a DataCard.
To access the Diagnostic Log, on the Main Menu select BIOMED, then
DIAG. LOG.
Diagnostic messages caused by an error condition can be generated by
four sources:
• Replaceable hardware malfunction, such as an etCO2 Cartridge or
HEMO Pod
• Software error
• Intermittent condition
• Hardware error
Notes for Table C-1:
1) When action says replace..., do so only if problem can be reproduced.
2) Convention of A, B, C etc. in the messages given in the Description
column is an abbreviation of actual messages, which will have
numbers in place of the A, B, C, etc., that help engineers uncover the
actual cause.
3) When action says "upgrade to Vxx", problem can also be solved by
upgrading to any SW version released after Vxx.
Description
Siemens Medical Systems, EM-PCS, Danvers
NOT A CONTROLLED DOCUMENT
SC 7000 and SC 9000XL Patient Monitors
Cause
Upgrade SW to ≥VC2.x.
Bad Laser Printer
Address reset
Try upgrading SW to
currently released version.
If error persists, replace
Main Processor Board
Subassembly.
Action
ASK-T898-03-7600
7k9kXLSM.apc.CD_ROM.fm/04-99/kaupp