SC 7000 and SC 9000XL Patient Monitors
Table C-1Diagnostic Log Codes (Continued)
Code
xxx = any
alphanumeric
characters
80271xxx
Network Communications - related
80272xxx
80273xxx
NIBP - related
80274xxx
8027dxxx
Network Communications - related
8027exxx
8027fxxx
80280xxx
ASK-T898-03-7600
7k9kXLSM.apc.CD_ROM.fm/04-99/kaupp
Description
Siemens Medical Systems, EM-PCS Danvers
NOT A CONTROLLED DOCUMENT
Cause
Try disconnecting/
reconnecting cable(s)
between Monitor and
network wallbox.
Verify proper SW version(s)
installed in all related
devices, in accordance with
SW compatibility chart for
Monitor SW.
If problem persists, try
swapping CPS/IDS with
known-good device.
If problem persists, replace
Main Processor Board
Subassembly.
Try disconnecting/
reconnecting pneumatic
tubing.
If problem persists, replace
Main Processor Board
Subassembly.
Try disconnecting/
reconnecting cable(s)
between Monitor and
network wallbox.
Verify proper SW version(s)
installed in all related
devices, in accordance with
SW compatibility chart for
Monitor SW.
If problem persists, try
swapping CPS/IDS with
known-good device.
If problem persists, replace
Main Processor Board
Subassembly.
Software did not
Try disabling items
allocate enough
connected or disabling Auto
memory for user's
Dual View.
current setup.
If error persists, replace
Main Processor Board
Subassembly.
Service Manual
Action
125