Table C-1Diagnostic Log Codes (Continued)
Code
xxx = any
alphanumeric
characters
8035dxxx
Network Communications - related
8035exxx
8036fxxx
80370xxx
80371xxx
MGM - related
80372xxx
130
Description
Siemens Medical Systems, EM-PCS, Danvers
NOT A CONTROLLED DOCUMENT
SC 7000 and SC 9000XL Patient Monitors
Cause
Try disconnecting/
reconnecting cable(s)
between Monitor and
network wallbox.
Verify proper SW version(s)
installed in all related
devices, in accordance with
SW compatibility chart for
Monitor SW.
If problem persists, try
swapping CPS/IDS with
known-good device.
If problem persists, replace
Main Processor Board
Subassembly.
Verify proper SW version(s)
installed in all related
devices, in accordance with
SW compatibility chart for
Monitor SW.
If problem persists, try
disconnecting/reconnecting
cable(s) between Monitor
and MGM.
If problem persists,try
replacing cable(s) between
Monitor and MGM.
If problem persists, try
swapping MGM with known-
good device.
If problem persists, try
swapping CPS/IDS with
known-good device.
If problem persists, replace
Main Processor Board
Subassembly.
7k9kXLSM.apc.CD_ROM.fm/04-99/kaupp
Action
ASK-T898-03-7600