Siemens SC 7000 Service Manual page 20

Patient monitors
Table of Contents

Advertisement

Service Manual
10.2 Service Menu
10.3 Install Monitoring
Software
4
The Service Menu is accessed via the Monitor Options selection under the
Monitor Setup function on the Main Menu. To access the Service menu and
related functions, do the following:
1) Select Monitor Setup on Level 1, then select Biomed on Level 2, and
then select Service on Level 3.
2) Input the service password (4712).
Note: In general, the Service Menu provides access to the following
(may vary with software version):
• Language selection
• Regulation
• Alarm Sounds
• Network control
• Network Configuration (requires installed SW version ≥VC2)
• Transport Brightness
• Line frequency setting
• Restore factory defaults
• Copy setups to card
• Copy setups to monitor
• Install Software
• Locked Options
• Waveform Simulator
Software and languages for SC 7000 and SC 9000XL Monitors are installed
from a memory card via the monitor's memory card reader. If the software
loading process fails to complete properly, and/or the monitor sounds a
steady tone (other than the Piezo), repeat the procedure. If the process fails
a second time, either the card or the Monitor is defective. Troubleshoot and
repair or replace as necessary.
Note: In an I
NFINITY
with <VC2 software, or ≥VC2 software and configured for CPS/IDS
mode, be sure that the CPS or IDS is equipped with compatibile
software. Refer to the Software Compatibility Chart for the CPS/IDS
software version required for compatibility with a specific SC 7000 or
SC 9000XL software version
1) With Monitor switched off, insert and firmly seat PCMCIA card into
memory card slot. Do NOT remove PCMCIA card until instructed to do
so.
Note: The card can be seated in only one orientation because of
keyed channels on the end of the card. If the card can not be easily
seated, remove card, turn card over, reinsert, and firmly seat. Do NOT
attempt to forceably seat the card.
2) Power Monitor ON to initiate download process.
Note: During the download process, the pick and go icon (running
man) and the SIEMENS logo appear on the screen. The icon initially
displays as green and changes to white. The logo toggles between
green and white,and finally displays as green on a white background.
The newly installed software version appears under the logo.
3) After a single alert tone sounds and a message regarding patient data
loss appears, select "Continue" and then select YES for new patient.
Siemens Medical Systems, EM-PCS, Danvers
NOT A CONTROLLED DOCUMENT
SC 7000 and SC 9000XL Patient Monitors
N
environment, if the monitor is equipped
ETWORK
7k9kXLSM.c1.CD_ROM.fm/04-99/kaupp
ASK-T898-03-7600