SC 7000 and SC 9000XL Patient Monitors
2.5 MPC 860
Communication
Channels
Main Processor
Front End Processor
2.6 Device CPS
2.7 Infinity Docking
Station
2.8 etCO
cartridge
2
2.9 External Battery
2.10 Interfaces
2.10.1 Local Fixed Keys Interface
2.10.2 Local Rotary Knob Interface
ASK-T898-03-7600
7k9kXLTM.c2.CD_ROM.fm/04-99/kaupp
MPC 860 has an embedded communications processor capable of
executing several protocols such as UART or Ethernet. The 860
communications channels are used as follows:
SCC1
Ethernet
SCC2
SC 9015 UART
SCC3
MVP-1 UART
SCC4
MVP-2 UART
SMC1
main diag UART
SPI
a/d (power monitor)
SCC3
serial pod data
SCC4
serial pod data
SMC1
front end diag UART
An additional UART implemented in the FPGA contains a large FIFO and
interfaces to the chart recorder.
A CPS (communication, power supply) has three boards -- power supply,
network board, and connector board, and allows for connection to an
INFINITY network as well as MIB and CAN. The FPGA and main program
memory are downloadable by the SC 7000 and SC 9000XL. The power
supply is a medical grade universal supply operating from 88 to 264 vac.
The Docking Station is a passive device that serves as a sturdy mount for
SC 7000 and SC 9000XL (as well as SC 9000 and SC 6000 series patient
monitors), in support of the P
through signals between the monitor and the CPS.
The Infinity Docking Station base unit (IDS) with its separate power supply,
replaces the CPS and Docking Station combination in an I
N
. As a docking station, it serves as a secure mount for the patient
ETWORK
monitors, as above, and also provides the network and peripheral device
connectivity of a Basic CPS. When equipped with a MIB 1-2 Option, it
provides the functionality of a Device CPS and Breakout Box.
The etCO
cartridge is a hardware component attached to the back of the
2
SC7000 or SC 9000XL. The cartridge is a self contained computer capable
of mainstream and side stream measurements.
The external battery is a low cost lead acid battery and may be charged in
either the monitor or with an external charger. The external battery may be
hot-swapped to extend battery life during transport.
There are twelve fixed function keys on the monitor base unit, and a fixed
key dedicated as a power on/off switch. The power on/off switch is unique
in that it is not directly available via a status read command. The on/off
switch is input to the power supply subsystem interface, where the switch
state is detected and processed. Detection of a power off condition causes
an interrupt to the host processor.
The rotary knob is a 16 detent rotary knob. Each detent position indicates
a "click" clockwise or counter-clockwise. The change in detent position is
detected via a 2 bit quadrature code that changes value every time the
rotary knob is moved into a detent position. Also included in the rotary knob
is a push button switch that is operated by a press/release action. This
switch is used to select menu items on the screen.
Siemens Medical Systems, EM-PCS Danvers
NOT A CONTROLLED DOCUMENT
10 Mbits/sec (future option)
selectable baud
selectable baud
selectable baud
19.2 Kbaud
-
-G
concept, and provides pass-
ICK
AND
O
Service Manual
NFINITY