12HemoMed Pod
12.1 IBP Function
12.1.1 IBP Test setup
68
300
290
280
270
260
250
240
230
220
210
200
190
180
170
160
150
140
130
120
110
100
90
80
70
60
5
40
20
0
Manometer
Pressure
Transducer
Tester
Figure 4-2 IBP Functional Verification Test Setup for HemoMed Pod
Note: As CO
is heavier than room air, set up the airway adapter
2
such that the point where the gas exits from the adapter is higher in
elevation than the point where it enters.
9) Place sensor on adult airway adapter and note reading = 0 ±1 mmHg.
10) Turn valve on the cylinder until it is fully open.
11) Wait for 30 seconds and record displayed value.
12) Close valve and remove sensor from airway.
13) Verify that measured value is in range (0.05 x local pressure) ±3,
rounded to nearest integer.
14) Select etCO
parameter box and set Atmospheric Pressure to AUTO.
2
15) Repeat steps 4 through 12.
16) Verify that measured value = previously measured value ±3, rounded
to nearest integer.
Refer to Figure 4-2 for test setup. Invasive blood pressure and cardiac
output functions of HEMO2/4 PODs are incorporated in the HemoMed Pod.
A single cable connects the Pod to the HemoMed 1input on the left side of
the SC 7000 or SC 9000XL. Pressure labels for HemoMed Pod channels
are set in the monitor.
1) With MultiMed cable and all other patient inputs unplugged from
monitor, power-cycle monitor. Select NO to clear display.
2) Plug cable from HemoMed Pod output into HemoMed 1 connector on
SC 7000 or SC 9000XL monitor. See Figure 4-2.
Siemens Medical Systems, EM-PCS, Danvers
NOT A CONTROLLED DOCUMENT
SC 7000 and SC 9000XL Patient Monitors
Inflation
Bulb
Stopcock
(Closed)
7k9kXLSM.c4.CD_ROM.fm/04-99/kaupp
ASK-T898-03-7600