Rear Housing Subassembly
Main Processor PCB Subassembly
Front Bezel Subassembly
Figure 3-9
Main Subassemblies of SC 7000 and SC 9000XL Patient Monitors (left side panel not shown). Note:
Numbers in black circles relate to corresponding numbers in white circles. See text.
50
5
5
3
1
as a specific replacement for a defective housing or Connector I/O
PCB. The required serial number is embedded in the new Connector
I/O PCB installed in the replacement subassembly, and the proper
label is placed on the back of the housing. It is the responsibility of
the individual replacing a defective Rear Housing Subassembly to
ensure destruction and proper disposal of the subassembly replaced.
1) Open monitor. Transfer Main Processor PCB Subassembly and Funnel.
2) Transfer NP Subassembly.
Note: Be sure to plug the NBP cuff connector sensor ring on new
housing into X8 on Connector I/O PCB. See
3) Close monitor and transfer remaining components and subassemblies
into the compartments on back of monitor under rear cover.
4) Functionally verify proper operation of reassembled monitor.
Siemens Medical Systems, EM-PCS, Danvers
NOT A CONTROLLED DOCUMENT
SC 7000 and SC 9000XL Patient Monitors
2
2
4
Figure 3-1 on page
ASK-T898-03-7600
7k9kXLSM.c3.CD_ROM.fm/04-99/kaupp
40.