14Memory Backup
Function
15CPS/IDS Mode
16DirectNet Mode
(requires ≥VC2-level
installed software)
72
3) Press C.O. START key on simulator, and verify an Injectate
Temperature indication of 1±0.2°C.
The monitor retains patient-related data, such as alarm limits, trends, and
stored alarm recordings when it is powered off.
1) With monitor powered-up and no patient inputs applied, press Alarms
Limits fixed key and change limit for any parameter.
Note: Note time on Real Time Clock.
2) Power monitor off for approximately 2 minutes, and then power it back
on again.
3) After MAIN screen displays, press Alarm Limits fixed key to call up
Alarm Limits Table.
4) Verify that new limit(s) you set in Step 1 have been retained, and that
clock has advanced the proper amount of time.
This procedure tests that monitor communicates with a CPS or IDS. Omit
this test and go on to Section 16 if monitor used exclusively in DirectNet
mode.
1) With monitor mounted on docking station connected to a CPS, or on
IDS press Menu key, and select Monitor Setup → Biomed → Service.
2) Enter biomed password (4712) and select "Accept."
3) Select Network Setup → Network Config. → CPS/IDS → Save ALL.
Note: Monitor resets if Network Config. previously set for DirectNet.
4) Press Menu key and select Monitor Setup → Biomed → Logs
5) Select Component Log → CPS/IDS.
6) Verify that Serial # of CPS or IDS appears on display.
This checks for proper configuration of a monitor set up for DirectNet
functioning, and for monitor communication via the I
this test if monitor used exclusively in CPS/IDS mode.
1) With monitor disconnected from docking station, connect monitor to
I
N
using ethernet connector on right side of monitor.
NFINITY
ETWORK
Note: Remove label covering ethernet connector on right-hand side of
monitor, if necessary.
2) Press Menu key, and select Monitor Setup → Biomed → Service.
3) Enter biomed password (4712) and select "Accept."
4) Select Network Setup → Network Config. → DirectNet → Save ALL.
Note: Monitor resets if Network Config. previously set for CPS/IDS.
5) Setup monitor for simulated patent using external patient simulator.
6) Verify that waveforms display on MVWS.
Note: If I
N
NFINITY
ETWORK
function of another monitor on network to view waveforms. Then skip
step 7 and go to step 8.
7) Access netInfo on MVWS and verify that monitor labels are unique.
8) Follow procedure in steps 2 through 4 to reset monitor to CPS/IDS
mode, if required.
Siemens Medical Systems, EM-PCS, Danvers
NOT A CONTROLLED DOCUMENT
SC 7000 and SC 9000XL Patient Monitors
NFINITY
not equipped with MVWS, use remote view
7k9kXLSM.c4.CD_ROM.fm/04-99/kaupp
N
. Omit
ETWORK
ASK-T898-03-7600