3) Verify that current does not exceed values given in Table 4-3.
4) Record all values in monitor's functional verification checklist.
1) With a partially discharged main (internal) battery and no external
(auxiliary) battery installed, place monitor on powered docking station
and bring up MAIN screen.
2) Slide lever on IDS to unlock position and note battery charge level.
3) Slide lever on IDS back to power position, and allow battery to charge
for ≈1 hr.
4) After 1 hr., verify that charge level on battery has increased.
5) With main battery at full charge and monitor on battery power, insert
external battery. Verify that battery charge level indicator at bottom of
MAIN screen changes from internal indicator to external indicator.
The R50 Recorder connects to the monitor through the CPS or IDS, or
through an interface plate. The following procedure verifies that the monitor
is communicating with the Recorder.
1) Connect known good R50 Recorder to monitor, either through a CPS/
IDS or via an interface plate.
2) Press Menu key, and access Monitor Setup → Biomed → Service
3) Enter Service password (4712) and select Accept.
4) Select Bedside Setup, and set Waveform Simulator to ON.
5) After monitor cycles, bring up MAIN screen.
6) Press Record fixed key.
7) Verify that recorder begins to dispense waveform printout.
8) Press Stop key on recorder to stop paper flow.
9) Press Menu key, and access Monitor Setup → Biomed → Service.
10) Enter Service password (4712) and select Accept.
11) Select Bedside Setup, and set Waveform Simulator to OFF.
Siemens Medical Systems, EM-PCS, Danvers
NOT A CONTROLLED DOCUMENT
SC 7000 and SC 9000XL Patient Monitors
ASK-T898-03-7600
7k9kXLSM.c4.CD_ROM.fm/04-99/kaupp