Siemens SC 7000 Service Manual page 127

Patient monitors
Table of Contents

Advertisement

SC 7000 and SC 9000XL Patient Monitors
18HEMO Pod
Pressure Front End Interface Connector
Figure B-21HEMO Pod Connectors (Refer to Table B-18.)
Table B-18 HEMO Pod Connector Signals
Pin No. PRESSURE FRONT END
INTERFACE
1
+SIG (P1)
2
GND
3
GND
4
+EXCIT (P1)
5
+EXCIT (P2)
6
GND
7
GND
8
+SIG (P2)
9
-SIG (P1)
10
GND
11
-EXCIT (P1)
12
GND
13
-EXCIT (P2)
14
GND
15
-SIG (P2)
ASK-T898-03-7600
7k9kXLSM.apb.CD_ROM.fm/04-99/kaupp
Temp A
TEMP 1
GND
GND
GND
GND
GND
GND
3
1
2
4
Figure B-22Press Adapter Input (Refer to Table B-19.)
Table B-19 PRESS Input Connector
Pin No. A, B, C, D
1
+EXCIT
2
3
-EXCIT
4
Cable In
5
-SIGNAL
Siemens Medical Systems, EM-PCS Danvers
NOT A CONTROLLED DOCUMENT
Temp A
C.O./Temp B
C.O./Temp B IBP SC 7000 / 9000XL
TEMP 2
GND
.7R37
TBLD
TINJ
GND
7
5
9
0
6
8
Pin No.
A, B, C, D
6
+SIGNAL
7
8
GND
9
0
Service Manual
IBP SC 7000
/ 9000XL
CNTRLH
CNTRLL
DATH
VDCRTN
VDC
DATL
111