Service Manual
Table C-1Diagnostic Log Codes (Continued)
Code
xxx = any
alphanumeric
characters
80117xxx
Serial Interface to Remote Keypad -
related
80118xxx
80119xxx
Temperature - related (MultiMed Pod)
8011axxx
80127xxx
bkg_operational determined
MONITORING SW is corrupt.
80128xxx
80129014
boot_program_CPS_images failed with
status: A on 68302 image.
8012901e
boot_validate_download_card could not
open/close one of the boot files.
8014bxxx
RECORDINGS - related
8014cxxx
118
Description
Siemens Medical Systems, EM-PCS, Danvers
NOT A CONTROLLED DOCUMENT
SC 7000 and SC 9000XL Patient Monitors
Cause
Try disconnecting/
reconnecting keypad.
If problem persists, swap
cable with known-good
keypad.
If problem persists, replace
Main Processor Board
Subassembly.
Try disconnecting/
reconnecting cable.
If problem persists, swap
cable with known-good
cable.
If problem persists try
swapping pod.
If problem persists, replace
Main Processor Board
Subassembly.
Background entity
Try upgrading SW to
determined that the
currently released version.
Main Software was
If error persists, replace
corrupted (BAD
Main Processor Board
CRC)
Subassembly.
CPS or IDS flash
Re-download CPS/IDS
memory could not be
software. If problem persists
programmed with
replace CPS or IDS.
new software
Software memory
Re-download monitor
card may have been
software. If problem persists,
removed during a
replace memory card.
download or the card
may have become
corrupted
Try disconnecting/
reconnecting cable.
If problem persists, swap
cable with known-good
cable.
If problem persists try
swapping Recorder.
If problem persists, replace
Main Processor Board
Subassembly.
Action
ASK-T898-03-7600
7k9kXLSM.apc.CD_ROM.fm/04-99/kaupp