SC 7000 and SC 9000XL Patient Monitors
10.4 Configuration
Download Procedure
ASK-T898-03-7600
7k9kXLSM.c1.CD_ROM.fm/04-99/kaupp
4) Access Bedside Setup, and verify that settings of Language,
Regulation, Alarm Sounds, Transport Brightness, and Line Frequency
are approrpriate for customer site. Also, assure that Waveform
Simulator is set to OFF.
5) Remove PCMCIA card.
6)• If SW version <VC2, or ≥VC2 and monitor to be mounted on a docking
station or IDS, set Monitor on active docking station and then set
Network Mode for CPS/IDS. Select "Save ALL" and then go to step 7.
Note: Monitor power-cycles automatically when "Save ALL" is
selected, if Network Mode is changed.
• If SW version ≥VC2, and monitor to be directly connected to I
N
, refer to procedure in Software Installation Instructions or
ETWORK
Service Setup Instructions to set Network Mode for DirectNet and
configure monitor. Then go to step 9.
7) Affix new software version label (supplied) over existing software
version label near top of memory card slot on right side panel.
8) Verify that monitor returns to MAIN screen, after timeout.
9) Recycle PCMCIA card when it is of no further use.
The configuration download procedure (-- not to be confused with monitor
configuration procedure required for DirectNet functioning; see
E: Service Setup
Instructions") is applicable only to monitors set for use
with a docking station or IDS. In general, the procedure is to completely set
up one monitor, save the setup to the CPS or IDS, and then transfer the
setup to a Data Card. The configuration stored in the Data Card can then
be used to setup monitor configurations in other CPS/IDSs.
1) With no Data Card inserted and monitor on docking station, adjust
settings for monitor exactly as required by customer.
2) Review configuration with appropriate customer personnel before
proceeding.
3) Press Menu key, and select Save/Restore → Save Setup.
4) Enter clinical password, 375, and select Accept.
5) Highlight setup to be saved, and press rotary knob in to save selection
to CPS/IDS.
6) Only the Default configuration supports "Pick and Go."
7) Wait for message "New Setup Saved."
8) Repeat steps
1
through 7 for optional setups as required, and select
Rename Setup in Biomed menu to name each setup in accordance
with site requirements.
9) With MAIN screen displayed on monitor, insert and firmly seat Data
Card into memory slot.
Note: The card can be fully inserted in only one orientation, because
of keyed channels on the end of the card, and can be damaged if
forced into the slot. Insert the card firmly, but do NOT attempt to
force the card. Be sure that Write Protect on the card is OFF.
10) Press Menu key, and select Monitor Setup → Biomed → Service.
Siemens Medical Systems, EM-PCS Danvers
NOT A CONTROLLED DOCUMENT
Service Manual
NFINITY