SC 7000 and SC 9000XL Patient Monitors
Table C-1Diagnostic Log Codes (Continued)
Code
xxx = any
alphanumeric
characters
8016a00a
Image checksum test failed due to TAXI
error!
80179xxx
8017axxx
8017fxxx
Temperature -related (MultiMed Pod)
80180xxx
8018301f
Unable to update CPS Chassis with
pick_n_go_mode
80185008
80187020
Error Log was cleared!
8018bxxx
etCO2 Pod/Cartridge - related
8018cxxx
8018d005
sn A, status B
ASK-T898-03-7600
7k9kXLSM.apc.CD_ROM.fm/04-99/kaupp
Description
Siemens Medical Systems, EM-PCS Danvers
NOT A CONTROLLED DOCUMENT
Cause
Bad memory/access/
Restart monitor. If error
fetch
persists replace monitor.
Pod communications
Try disconnecting/
problem.
reconnecting cable.
Pod cable, Pod, or
If problem persists, swap
CPU board failed
cable with known-good
cable.
If problem persists try
swapping pod.
If problem persists, replace
Main Processor Board
Subassembly
Try disconnecting/
reconnecting cable.
If problem persists, swap
cable with known-good
cable.
If problem persists try
swapping pod.
If problem persists, replace
Main Processor Board
Subassembly
Write failed to CPS or
replace CPS or IDS
IDS chassis
EEPROMs
Upgrade SW to ≥VC1.x.
Pick and Go reset
Manual clear of error
None
log via system
console menu
Try swapping out airway
adaptor with known-good
components.
Try disconnecting/
reconnecting cable.
If problem persists, swap
cable with known-good
cable.
If problem persists try
swapping pod.
If problem persists, replace
Main Processor Board
Subassembly.
etCO
Hardware Fail
Replace etCO
2
Service Manual
Action
cartridge.
2
119