SC 7000 and SC 9000XL Patient Monitors
Table C-1Diagnostic Log Codes (Continued)
Code
xxx = any
alphanumeric
characters
80373xxx
MIB - related
80374xxx
80375xxx
ISD (SDC) - related
80376xxx
ASK-T898-03-7600
7k9kXLSM.apc.CD_ROM.fm/04-99/kaupp
Description
Siemens Medical Systems, EM-PCS Danvers
NOT A CONTROLLED DOCUMENT
Cause
Verify proper SW version(s)
installed in all related
devices, in accordance with
SW compatibility chart for
Monitor SW.
If problem persists, try
disconnecting/reconnecting
cable(s) between Monitor
and MIB converter.
If problem persists,try
replacing cable(s) between
Monitor and MIB converter.
If problem persists, try
swapping MIB converter with
known-good device.
If problem persists, try
swapping CPS/IDS with
known-good device.
If problem persists, replace
Main Processor Board
Subassembly.
Verify proper SW version(s)
installed in all related
devices, in accordance with
SW compatibility chart for
Monitor SW.
If problem persists, try
disconnecting/reconnecting
cable(s) between Monitor
and ISD (SDC).
If problem persists,try
replacing cable(s) between
Monitor and ISD (SDC).
If problem persists, try
swapping ISD (SDC) with
known-good device.
If problem persists, try
swapping CPS/IDS with
known-good device.
If problem persists, replace
Main Processor Board
Subassembly.
Service Manual