SC 7000 and SC 9000XL Patient Monitors
Table C-1Diagnostic Log Codes (Continued)
Code
xxx = any
alphanumeric
characters
8036f000
8038d001
803c1000
803f1xxx
etCO2 Pod/Cartridge - related
803f2xxx
803f3xxx
803f4xxx
ffffffff
boot_program_images failed with
status: A on image: B
ASK-T898-03-7600
7k9kXLSM.apc.CD_ROM.fm/04-99/kaupp
Description
Siemens Medical Systems, EM-PCS Danvers
NOT A CONTROLLED DOCUMENT
Cause
Upgrade SW to ≥VC1.x.
Bad Network Config
Data Reset
Upgrade SW to ≥VC2.x.
MIB Connection/
Disconnection reset
Upgrade SW to ≥VC1.x.
Remote view, auto
dual view reset
Try swapping out airway
adaptor with known-good
components.
Try disconnecting/
reconnecting cable.
If problem persists, swap
cable with known-good
cable.
If problem persists try
swapping pod.
If problem persists, replace
Main Processor Board
Subassembly.
Flash programming
Retry download. If problem
error when
persists replace Main
performing a
Processor Board
download
Subassembly if downloading
monitor SW. replace CPS or
IDS, If downloading CPS/
IDS software.
Service Manual
Action