Service Manual
Table C-1Diagnostic Log Codes (Continued)
Code
xxx = any
alphanumeric
characters
8020fxxx
80210xxx
80231002
LCOM_BAD_PRL_SUPPLIED_ERROR
80231005
LCOM_BAD_SWITCH_COUNT_ERRO
R
8023100a
LCOM_CHECKSUM_DIAG_ERROR
80231025
LCOM_PRL_ENTRY_DIAG_ERROR
80231026
LCOM_PRL_ID_DIAG_ERROR
8023102b
LCOM_RAM_DIAG_ERROR
122
Description
Siemens Medical Systems, EM-PCS, Danvers
NOT A CONTROLLED DOCUMENT
SC 7000 and SC 9000XL Patient Monitors
Cause
Software did not
Try disabling items
allocate enough
connected or disabling Auto
memory for user's
Dual View.
current setup.
If error persists, try
upgrading SW to currently
released version.
If error persists, replace
Main Processor Board
Subassembly.
SW or comm error
Check cable connection. If
between CPS or IDS
no second error, replace
and Monitor
CPS or IDS.
Comm error
If this message is
accompanied by another
error, check Docking station
connections. Otherwise this
indicates a SW error and
should be reported to
Customer service via
DataCard.
Comm error
If this message is
accompanied by another
error, then check Docking
station connections.
Otherwise this indicates a
SW error and should be
reported to Customer
service via DataCard.
CPS or IDS harware
Reboot monitor and CPS or
failure
IDS. If problem persists,
replace CPS or IDS.
CPS or IDS hardware
Reboot monitor and CPS or
failure.
IDS. If problem persists,
replace CPS or IDS.
Net comm failure
If this message is
accompanied by another
error, check Docking station
connections. Otherwise this
indicates a SW error and
should be reported to
Customer service via
DataCard.
Action
ASK-T898-03-7600
7k9kXLSM.apc.CD_ROM.fm/04-99/kaupp