SC 7000 and SC 9000XL Patient Monitors
13etCO
Malfunction.
2
Table 5-10 etCO
Malfunctions
2
Symptom(s)
Parameter box fails to
appear when sensor
plugged into module
Sensor fails calibration
Persistent Adapter
Failure message
14HEMO2/4 Pod / HemoMed Pod
14.1 Readings Missing or
Inaccurate
Aux/Hemo2
Aux/Hemo3
ASK-T898-03-7600
7k9kXLSM.c5.CD_ROM.fm/04-99/kaupp
Possible Cause(s)
Sensor or cable
malfunction
etCO
module
2
malfunction
Connector I/O Board
malfunction
Main Processor PCB
malfunction
Sensor or cable
malfunction
etCO
module
2
malfunction
Main Processor PCB
malfunction
Airway adapter or sensor
window occluded
Airway adapter
malfunction
Sensor malfunction
Disruption in communications to an external pod is caused by the pod itself,
by a cable problem, or by a communication problem on the main processor
board. Power for an external pod is supplied from the main processor board
when the board senses that a load is present. If the monitor is not
reporting connection of a Hemo2/4 pod, 12-lead pod or tcpO2/CO2 pod, use
the following test to determine first whether or not the pod is being detected.
Note: Only the Hemo2/4 pod "communicates" with the monitor. The
Hemo2/4 Pod is a digital device, and the HemoMed Pod is an analog
or
device.
Connect a 150Ω ¼W resistor between pins 1 and 3 on the suspect port (see
illustration at left), and check monitor display for a pod connection error
"DEVICE FAILURE ON HEMO CONNECTOR x."
• If no error message displays, the problem is on the main processor
board. Replace the Main Processor PCB Subassembly.
• If error message displays, problem may be in cable, pod, or main
processor board. Replace defective component as required.
If only one input or output on a HEMO2/4 or HemoMed Pod fails to function,
try replacing the sensor or cable. If problem persists, replace the Pod.
Siemens Medical Systems, EM-PCS Danvers
NOT A CONTROLLED DOCUMENT
Troubleshooting and Remedial Action
1. Clean calibration windows.
2. Replace etCO
Sensor.
2
3. If problem persists, replace etCO
4. If problem persists, replace Main Processor PCB
Subassembly.
5. If problem persists, replace Rear Housing
Subassembly.
6. If problem persists, contact TSS in Solna or
Danvers.
1. Replace etCO
Sensor.
2
2. If problem persists, replace etCO
3. If problem persists, replace Main Processor PCB
Subassembly.
4. If problem persists, contact TSS in Solna or
Danvers.
1. If adapter or sensor window occluded, clean window.
2. If problem persists, replace airway adapter.
3. If problem persists, replace sensor.
4. If problem persists, replace etCO
Service Manual
Module.
2
Module.
2
Module.
2
87