Removal and Replacement Procedures

Connector Replacement Procedure

Connector Replacement Procedure

Parts Required

— SMA or 3.5 mm connectors
— Nuts for connectors

Tools Required

— "Spark plug" adapter
— Torque wrench (for "spark plug" adapter)
— ESD equipment:
Description
ESD grounding wrist strap
5-ft grounding cord for wrist strap
2 x 4 ft conductive table mat and 15-ft grounding wire
ESD heel strap (for use with conductive floors)

Removal Procedure

Although the photos in this section show a 3.5 mm connector being
replaced, the procedure is the same for SMA connectors.
1. Place the module on its side on an ESD safe, flat surface.
2. Place the "spark plug" adapter on the nut of the connector you are going to
Figure 5-1
Tools and Parts
Keysight P937xA PXIe Service Guide
replace. The adapter must fit snugly into the slots of the nut. Refer to
5-2 on page
64.
Keysight Part Number
9300-1367
9300-0980
9300-0797
9300-1308
Figure
63