4.) Press the ON/OFF Standby switch left below the Control Panel once.
ON/OFF Standby Switch
When the ON/OFF Standby switch left below the Control Panel is pressed once, the System (including
the Back-end Processor) starts and the software code is distributed to initiate the scanner.
The company logo appears on the Touch Panel during boot up.
Depending on the BIOS-Version no status messages are displayed during this process.
Boot up time is about 3 minutes.
NOTE:
The mains outlet of the system for peripheral auxiliary equipment are commonly switched with
the Standby switch. So the auxiliary equipment need not to be switched ON/OFF separately.
4-3-2
Power Off / Shutdown
NOTE:
After turning off a system, wait at least ten seconds before turning it on again. The system may not be
able to boot if power is recycled too quickly.
4-3-2-1
Scanner Shutdown
1.) Press the ON/OFF Standby switch left below the Control Panel once.
2.) Switch OFF the Circuit Breaker at the rear of the system.
3.) Disconnect the Mains Power Cable is necessary. For example: Relocating the scanner.
Circuit Breaker
Mains Power Cable
Figure 4-3 Circuit Breaker and Power Cable on Back of Voluson® 730
U
RETZTECHNIK
LTRASOUND
1
Figure 4-2 ON/OFF Stand-By Switch
Chapter 4 Functional Checks
V
® 730 S
OLUSON
ERVICE
M
ANUAL
4-3