GE Voluson 730 Service Manual page 188

Hide thumbs Also See for Voluson 730:
Table of Contents

Advertisement

GE M
S
- K
EDICAL
YSTEMS
D
105844, R
IRECTION
EVISION
CD-RW Troubleshooting (
7-7-7-1
1.) Insert an empty CD-RW into the Drive.
2.) Enter "Sonoview" by pressing the NETWORK key on the control panel; see
3.) Click the "Open" icon to display the list of exams.
4.) Select exam(s) and backup them to CD-ROM.
5.) Choose "CD-ROM" Drive.
6.) The images, which you have chosen during backup should be visible.
1. Press Network key
7-7-8
MOD Troubleshooting
1.) Insert an empty MO into the Drive.
2.) Enter "Sonoview" by pressing the NETWORK key on the control panel; see
3.) Click the "Open" icon to display the list of exams.
4.) Select exam(s) and backup them to MO.
5.) Choose "MO" Drive.
6.) The images, which you have chosen during backup should be visible.
7-7-9
Audio Test
Preparations for Audio Test: Start a probe and select the PW-Mode.
Change the volume from 0 to 96dB with the digipot and listen if both loudspeakers are making some
noise.
7-18
U
RETZTECHNIK
LTRASOUND
1
2. Click "Open" icon
and backup exams
Figure 7-20 To backup exams to CD-ROM resp. MO-Disk
Section 7-7 - Troubleshooting Trees and Instructions
CD-Rom Drive)
V
® 730 S
OLUSON
ERVICE
Figure
7-20.
3. After Backup,
choose CD-ROM
(resp. MO)
Figure
7-20.
M
ANUAL

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Voluson 730 and is the answer not in the manual?

Subscribe to Our Youtube Channel

Table of Contents