Save System Presets and Configurations (Application Settings)
The Backup contains:
User Settings, Auto Text, Measurement user tables
Biopsy (needle guide positions)
Setup settings (language, date format, screensaver on/off, etc.)
NOTE:
Always backup any preset configurations before upgrading the Application Settings.
This ensures that if the presets need to be reloaded, after the software update, will be the same ones
the customer was using prior to service.
Saving the Backup to CD-RW is impossible!
NOTICE
1.) Insert a formatted MOD (Magneto-Optical Disk) into the drive.
2.) On the Touch Panel, touch UTILITIES.
3.) On the Utilities Touch Panel, touch SYSTEM SETUP to invoke the setup desktop on the screen.
4.) Select the
5.) Choose the media (e.g.MO-Drive) and click the SAVE button.
6.) Select the NEW FILE... button and enter a file name (backup name - without extension).
7.) Click the OK button. The system performs the backup.
8.) When the saving of the System Settings has been completed, click OK.