Scan Probe Environmental Requirements; Time And Manpower Requirements - GE Voluson 730 Service Manual

Hide thumbs Also See for Voluson 730:
Table of Contents

Advertisement

GE M
S
- K
EDICAL
YSTEMS
D
105844, R
IRECTION
EVISION
2-2-4

Scan Probe Environmental Requirements

Operation:1Ambient temperature 18° to 30° C
Storage:-10° to 50° C
NOTE:
Temperature in degrees C. Conversion to degrees F = °C (9/5) + 32).
SYSTEMS AND ELECTRONIC PROBES ARE DESIGNED FOR STORAGE TEMPERATURES OF
NOTICE
-10 TO + 50 degrees C. WHEN EXPOSED TO LARGE TEMPERATURE VARIATIONS, THE
PRODUCT SHOULD BE KEPT IN ROOM TEMPERATURE FOR 10 HOURS BEFORE USE.
2-2-5

Time and Manpower Requirements

Site preparation takes time. Begin Pre-installation checks as soon as possible. If possible, allow six
weeks before delivery, for enough time to make necessary changes.
Have two people available to deliver and unpack the Voluson® 730.
CAUTION
Attempts to move the unit considerable distances or on an incline by one person could result
in injury or damage or both.
U
RETZTECHNIK
LTRASOUND
1
Chapter 2 Pre-Installation
V
® 730 S
OLUSON
ERVICE
M
ANUAL
2-5