8-3-2
EUM - Upgrade Procedure (from Software Version 1.06G)
8-3-2-1
EUM - Removal Procedure
1.) Restart the system (turn off and on the system).
The electronic user manual (EUM) must never be opened (by pressing the F1 key on the keyboard)
NOTICE
after last restart! Even if closed again before installing the new EUM.
2.) Touch the UTILITIES key on the Touch Panel once to display the Utilities menu.
3.) Touch the SYSTEM SETUP key to invoke the setup desktop on the screen.
4.) Select the
The "Service Tools" menu appears on the screen.
5.) Click the EUM button. The "Update Software" dialog appears.
6.) Select the MO DRIVE or the CD DRIVE button, depending on the storage medium you use.
7.) Browse for the EUMSetup_en_xxxx.exe file and click OK.
U
RETZTECHNIK
LTRASOUND
1
page, enter the password SHE, and click the ACCEPT button.
SERVICE
Figure 8-3 Service Tools
Figure 8-4 Browse for file
Chapter 8 Replacement Procedures
V
® 730 S
OLUSON
ERVICE
EUM button