GE Voluson 730 Service Manual page 92

Hide thumbs Also See for Voluson 730:
Table of Contents

Advertisement

GE M
S
- K
EDICAL
YSTEMS
D
105844, R
IRECTION
EVISION
4-4-1
2D-Mode Checks
Table 4-3
Step
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
4-8
U
RETZTECHNIK
LTRASOUND
1
(cont'd)
2D-Mode Functions
Task
2D-Mode Gain
Transmit Power
Focus Depth
Depth
Screen Format (Dual, Quad)
FFC (Focus and Frequency Composite)
CRI (Compound Resolution Imaging)
LINEAR / TRAPEZ
Image Orientation
ANGLE
ß-VIEW
ZOOM
FOC. ZONES
OTI (Otimized Tissue Imaging)
FREQUENCY
GRAY CHROMA MAP
Section 4-4 - Functional Checks
Expected Results
Rotate the 2D-MODE key to adjust the sensitivity (brightness) of the entire
image.
Optimizes image quality and allows user to reduce beam intensity.
To select the depth position of the actual focus zone(s). Arrows at the left
edge of the 2D-Image mark the active focal zone(s) by their depth position.
Adjusts the depth range of the ultrasound image for the region of interest.
The number of image lines and the frame rate are automatically optimized.
Press this keys to change the display Mode from Single to DUAL or QUAD
display mode.
Press the 2D-MODE key to change from Dual or Quad to Single display.
FFC combines a low frequency to increase the penetration and higher
frequency to keep a high resolution.
It reduces speckle and artifacts in the 2D-image.
Pulses are transmitted not only perpendicularly to the acoustic window, but
also in oblique directions. The advantages of CRI are enhanced contrast
resolution with better tissue differentiation and clear organ borders.
Advantage of the Trapezoid-Mode: The scan area is very increased in
relation to the linear display by steering the ultrasound lines in the border of
the probe.
Use the LEFT/RIGHT respectively the UP/DOWN keys on the Touch Panel
to alternate the image orientation.
Use this control to select a part of interest of the 2D-Image. The advantage
of the decreased field-of-view is an increased 2D frame rate due to the
smaller sector width.
This function allows the adjustment of the Volume O-Axis position of 3D
probes in 2D-Mode. The green line in the displayed symbol indicates the
position of the acoustic block.
Image magnification (Pan Zoom) in Read-/ and Write-Mode.
Increases the number of transmit focal zone, so that you can tighten up the
beam for a specific area.
OTI™ allows to "fine tune" the system for scanning different kinds of tissue.
To adjust the range of the receive frequency.
high resolution / lower penetration, mid resolution / mid penetration, or
lower resolution / high penetration
A gray map determines the displayed Brightness of an echo in relationship
to its amplitude.
V
® 730 S
OLUSON
ERVICE
M
ANUAL