Physical Inspection
.
Table 10-9
Step
1
2
3
4
5
6
Cables & Connectors
7
8
9
10
11
External Microphone Check for proper operation of any external microphones by recording an audio test if available.
There is no microphone built in and released for Voluson® 730.
NOTICE
U
RETZTECHNIK
LTRASOUND
1
Physical Checks
Item
Labeling
Verify that all system labeling is present and in readable condition.
Scratches & Dents
Inspect the console for dents, scratches or cracks.
Control Panel
Inspect keyboard and control panel. Note any damaged or missing items.
Control Panel
Verify ease of control panel (Operator I/O Panel) movement in acceptable direction. Ensure that
Movement
it latches in position as required.
Check all wheels and casters for wear and verify operation of foot brake, to stop the unit from
Wheels & Brakes
moving, and release mechanism. Check all wheel locks for proper operation.
Check all internal cable harnesses and connectors for wear and secure connector seating.
Pay special attention to footswitch assembly and probe strain or bend reliefs.
Check to ensure that all EMI shielding, internal covers, air flow panels and screws are in place.
Shielding & Covers
Missing covers and hardware could causeEMI/RFI problems while scanning.
External I/O
Check all connectors for damage and verify that the labeling is good.
Check for proper operation of all operator panel key illuminations (flash once during system start-
Op Panel Lights
up).
Monitor Light
Check for proper operation of any monitor lights if available.
Chapter 10 Periodic Maintenance