GE Voluson 730 Service Manual page 86

Hide thumbs Also See for Voluson 730:
Table of Contents

Advertisement

GE M
S
- K
EDICAL
YSTEMS
D
105844, R
IRECTION
EVISION
Section 4-3
General Procedure
SYSTEM REQUIRES ALL COVERS
CAUTION
Operate this unit only when all board covers and frame panels are securely in place. The
covers are required for safe operation, good system performance and cooling purposes.
Lockout/Tagout Requirements (For USA only)
NOTICE
Follow OSHA Lockout/Tagout requirements by ensuring you are in total control of the Power Cable
on the system.
TAG
&
LOCKOUT
Signed
Date
4-3-1
Power On / Boot Up
NOTE:
After turning off a system, wait at least ten seconds before turning it on again. The system may not be
able to boot if power is recycled too quickly.
4-3-1-1
Scanner Power On
1.) Connect the Power Cable to the back of the system.
2.) Connect the Power Cable to an appropriate mains power outlet.
3.) Switch ON the Circuit Breaker at the rear of the system.
Circuit Breaker
Mains Power Cable
Figure 4-1 Circuit Breaker and Power Cable on Back of Voluson® 730
When AC power is applied to the scanner, the ON/OFF switch on the Control panel illuminates amber,
NOTICE
indicating the System (including the Back-end Processor) is in standby mode.
4-2
U
RETZTECHNIK
LTRASOUND
1
Section 4-3 - General Procedure
V
® 730 S
OLUSON
ERVICE
M
ANUAL

Advertisement