D-Mode (Pulsed Wave- and Continuous Wave Doppler)
1.) CPC
see:
2.) CPR
see:
3.) CPG
After DC-cancelling the sig is mixed with the NCO-generated RX-Frequency and brought to LF-
Spectrum, where the FIR-LP cuts HF. Output to CPF.
4.) CPF
Arranges the FFT. D-Mode Data use the same Bus to the CPS as CFM-Data.
5.) CPS
D-Mode Data from CPF are written... see
6.) CKV
see:
5-2-4-4
CFM-Mode (Color Flow Mode)
1.) CPC
see:
2.) CPR
see:
3.) CPG
After DC-cancelling the sig is mixed with the NCO-generated RX-Frequency and brought to LF-
Spectrum, where the FIR-LP cuts HF. Output to CPF.
4.) CPF
Arranges the CFM-Processing.
WMF (Wall Motion Filter), Autocorrelator, Post-processing.
5.) CPS
Data from CPG and CPF are written... see
6.) CKV
see: