Setting Up The Kodak Process Control Densitometer - Kodak Ektascan 2180 User Manual

Hide thumbs Also See for Ektascan 2180:
Table of Contents

Advertisement

The following illustration represents the test pattern film showing the 21-step gray scale
on the right side of the film.

Setting Up the Kodak Process Control Densitometer

For complete instructions on setting up the Kodak Process Control Densitometer, refer
to the Operation Manual for the Kodak Process Control Densitometer, Publication Part
Number 635933, available from X-Rite (616) 534-7663.
Note
The Densitometer has 16 channels. Each channel monitors a single processor. Therefore,
it is necessary to dedicate and set up a channel for each processor in your department.
There are 4 parameters which should be monitored:
• Speed Index (S Index)
• Base+Fog (B+Fog)
• Customer Upper Density Point (CUDP)
• Developer Temperature (Temp °F)
Note
Contrast Index (C.Indx) does not need to be monitored.
It is recommended that the factory preset Densitometer value of "Dmax" be changed to
CUDP value to more accurately describe the variable monitored. For more information,
refer to the section "Determining the Customer Upper Density Point" on page 16–8.
5B9615
Using a 21-Step Gray Scale for Processor Quality Assurance
USE FOR CALIBRATION
1
2
3
4
5
6
7
8
9
General CT1 12:03:09 15-May-93
Rochester, NY
H126_0032AA
16–3

Advertisement

Table of Contents
loading