Flooded Replenishment Mode
In the Flooded Replenishment Mode, the preset values for developer and fixer
replenishment volumes are:
• 65 mL every 5 minutes.
• 65 mL for each 1500 cm
If you wish, you may change the setup information by using the function keys on the
display panel. See "Changing Setpoints" in this section.
Important
Before changing the replenishment volumes, calibrate the replenishment system.
See "Calibrating the Replenishment System" in this section.
Press [F2] at the Processor Setup Screen to display the Processor Replenishment Mode
Screen.
13–8
2
2
(238 in.
) of film.
PROCESSOR REPLENISHMENT MODE
Processor Ready
Replenishment Mode is Auto
F1
F2
Select Flooded
F3
Fill Tanks
F4
Disable Replenishment
F6
Return
Select the desired function.
H126_9053AC
5B9615