[8] Clean the Detector Crossover Assembly with a soft fiber brush and warm water.
Allow the Detector Crossover Assembly to air dry before processing film.
[9] Check that all Rollers on all Racks rotate freely.
[10] Check to be sure that the Racks and Crossover Assemblies are square by doing the
following procedure.
a. Place the Rack or Crossover Assembly on a smooth, flat surface.
b. If the Rack or Crossover Assembly is not level with the flat surface, press
down to make it level.
c. Check the squareness of the Side Plates.
d. If the Assembly still doesn't seem to be square, call your Service
Representative and ask that the proper adjustments be made.
[11] Install the Racks, Crossover Assemblies, and Wet Section Cover in the correct
positions in the Processor as described and illustrated in "Weekly" in this section,
and check that each Assembly seats firmly.
[12] Every 3 months calibrate the Replenisher Pumps. See "Calibrating the
Replenishment System" in Section 13.
[13] Check the Replenisher Strainers which are located between the Replenishment
Tanks and the Pumps, and clean the Strainers if necessary.
[14] Check the Water Supply Filter. Change the Filter if necessary.
5B9615
Preventive Maintenance
15–9