Maintaining the Densitometer
The Reference Strip is used to calibrate the Densitometer on the front of the 2180 Laser
Printer. The Densitometer is calibrated at installation time, but the Reference Strip can
be used periodically to check that the Densitometer is still properly calibrated.
A Reference Strip is delivered with each Printer. It can be found in the pocket on the
inside of the Film Door of the Printer. Replacement Reference Strips can be purchased
separately by ordering part number 748682.