Kodak Ektascan 2180 User Manual page 206

Hide thumbs Also See for Ektascan 2180:
Table of Contents

Advertisement

User's Manual
Warning
Wear rubber gloves, safety glasses, and protective clothing when doing any
maintenance on the 180 LP or 180 LPS Processor.
[4] Lift the Top Cover of the Processor.
Caution
Handle all Racks carefully to prevent changing their alignment or causing damage. Do
not clean the Racks, Crossover Assembly or Squeegee Rollers with abrasive materials.
[5] Remove the Wet Section Cover, all Crossover Assemblies, and the Wash Rack.
[6] Clean all Crossover Assemblies, the Wash Rack, and the Crossover Rack Guide
Shoes by wiping them with a soft, damp cloth and rinsing them with clean, warm
water.
[7] Clean the Detector Crossover Assembly with a soft fiber brush and warm water.
Allow the Detector Crossover Assembly to air dry before processing film.
[8] Check that all Rollers on the Wash Rack and the Crossover Assemblies rotate
freely.
[9] Check that the Wash Rack and Crossover Assemblies are square by doing the
following:
• Place the Rack or Crossover Assembly on a smooth, flat surface.
• If the Rack or Crossover Assembly is not level with the flat surface, press
down to make it level.
• Check the squareness of the Side Plates.
• If the Assembly still is not square, call your Service Representative and ask
that the proper adjustments be made.
15–4
5B9615

Advertisement

Table of Contents
loading

This manual is also suitable for:

X-omat 180 lpX-omat 180 lps

Table of Contents