[3] Process the film and check that the information on the screen matches the
destination printed on the film and the film type. If the information does not match,
use [F1] and [F2] to select the correct values. The Calibration Test Film will look
like the following illustration.
General CT1 12:03:06 15-May-92
Rochester, NY
calibration test film
[4] Do either step 5 or step 6 to enter the density data into the Printer. In step 5, the
density data is read by the built-in Densitometer and entered into the software of
the Printer. Do step 6 if you measure the density values at an off-line Densitometer
and then enter them at the Control Panel.
Note
Do not place the Calibration Test Film in the slot of the Densitometer until
prompted to do so.
5B9615
Calibrating the 2180 Laser Printer
USE FOR CALIBRATION
1
2
3
4
5
6
7
8
9
H126_2722AC
6–7