Kodak Ektascan 2180 User Manual page 125

Hide thumbs Also See for Ektascan 2180:
Table of Contents

Advertisement

Calibration Data Entry Screen
You may use another Densitometer instead of the built-in Densitometer that is on the
front of the 2180 Laser Printer to measure the density values of the Calibration Test Film.
If so, you should follow these steps to enter the density data into the Printer at the Control
Panel.
Press [F5] at the Calibrate Printer Screen to display the Calibration Data Entry Screen.
[1] Press [F1] to select the previous density step, moving through the list in a circular
manner.
Press [F2] to select the next density step.
Press [F3] to increase the selected value in the range of 0.00 to 3.99. The key may
be held down in autorepeat mode.
Press [F4] to decrease the selected value in the range of 0.00 to 3.99. Repeat this
process as many times as is necessary, and then continue to step 2.
[2] Press [F5] to compute calibration. If calibration is not successful, the Error
Recovery Screen will be displayed. If the calibration is successful, the Printer
changes the prompt to "Printer calibration successful."
[3] Press [F6] to return to the previous screen.
5B9615
CALIBRATION DATA ENTRY
1)
0.25
4)
1.31
2)
0.32
5)
1.95
3)
0.69
6)
2.40
F1
Previous Step
F2
Next Step
F3
Increase
F4
Decrease
F5
Compute Calibration
F6
Return
Select density steps and modify;
when all done, Compute.
Calibrating the 2180 Laser Printer
Ready
7)
2.71
8)
2.96
9)
3.21
H126_9044AC
6–9