Kodak Ektascan 2180 User Manual page 171

Hide thumbs Also See for Ektascan 2180:
Table of Contents

Advertisement

Processor Overview
Processor Error Messages are listed and described in Section 18 on page 18–15.
If the Processor Status line does not show that the Processor is off-line, then the
Developer, Fixer, and Dryer actual temperatures and temperature setpoints are
displayed. Decimal points or commas and temperature scale (Fahrenheit or Celsius) are
selected at the time of the configuration of your Printer.
Press [F1] to increase the dryer setpoint in increments of 1.0° Fahrenheit
(or 1.0° Celsius, depending on which temperature scale was selected for your site).
Press [F2] to decrease the dryer setpoint in increments of 1.0° Fahrenheit
(or 1.0° Celsius).
Press [F3] to display the Processor Setup Access Screen. This selection only appears as
an option on the screen when there is no film in the Processor.
Press [F4] to continue even though an error condition relating to temperature warnings
exists. If the Error description field is blank, or if it describes an error that cannot be
overridden, this option will not be available.
Press [F6] to return to the Other Functions Screen.
5B9615