User's Manual
error description field
F1
commands
The Processor Status field indicates the current condition of the Processor. These
conditions are "Processor Ready," "Processor Not Ready," "Film in Processor," or
"Processor Off-line."
The error description field provides an error number of any error that may occur in the
Processor, such as, "Processor Error: E134." When errors are initially reported, the Error
Recovery Screen displays the error text. After returning to this screen, just the numeric
code will appear as illustrated above. If this error is a Processor error that the user
chooses to override, this line will be displayed in flashing mode. If there is not an error,
this line will be blank. There are 3 errors that you can override. They are "E132
Developer under temperature," "E133 Developer over temperature," and "E134 Dryer
under set temperature."
11–4
Processor status field
CONTROL PROCESSOR
Processor Ready
Processor Error: Exxx
Dev:
temp=100.0F
Fix:
temp=100.0F
Dry:
temp=119.5F
F1
Increase Dry Set
F2
Decrease Dry Set
F3
Setup Processor
F4
Override Error
F6
Return
Select the desired function.
F2
F3
F4
prompt lines
Set=100.0F
Set=100.0F
Set=120.0F
F5
F6
H126_9048GCA