Changing the Encoding Screws
To change the Supply Magazine Encoding Screws, you will need a 2.5 mm Hex head
wrench.
On the bottom of the Supply Magazine, there are Encoding Screws which are sensed by
Switches to communicate information about the film type to the Printer. With the
Magazine upside down and the Handle to the left, locate Encoding Screws X1, X2 and
X3 as illustrated below.
The positions of Screws X1 and X2 define film size. The position of Screw X3 defines
the film base type.
Note that in the above illustration, there is a Screw in X2 but not in X1 or X3, indicating
that this Supply Magazine is set up for 11 x 14 in. blue base film.
Film size
X1
8 x 10 in.
screw
11 x 14 in.
no screw
35 x 35 cm screw
35 x 43 cm no screw
5B9615
Film base
X2
Clear
screw
Blue
screw
no screw
no screw
X3
X2
X1
H126_6331AA
X3
screw
no screw