Calibrating the Replenishment System

Procedure
Important
Calibrate the replenisher pumps every 3 months.
Calibration of the replenishment system will determine the actual rate of solution flow
through the Replenishment Pumps for a given period of time.
Measure the volume of solution that is delivered by the system, then use the Display
Panel to enter the measurement into the microprocessor. The microprocessor computes
the rate of solution flow through the Pump, then adjusts the period of time that the Pump
must operate to deliver the replenishment volume that was set.
Note
The volume actually measured during this procedure is not the volume that is delivered
for a 35 x 43 cm sheet of film during normal equipment operation.
[1] To calibrate the replenishment system, press [F4] at the Other Functions Screen to
see the Control Processor Screen.
[2] Press [F3] at the Control Processor Screen to display the Processor Setup Access
Screen.
[3] Enter the four-digit Access Code: 4213.
13–10
PROCESSOR SETUP
Processor Ready
Processor Error: E134
Cycle is Rapid
Replenishment Mode is Auto
F1
F2
Change Replen Mode
F3
Setup Process
F4
Dev Replen Calibration
F5
Fix Replen Calibration
F6
Return
Select the desired action.
H126_9055AC