Resetting The Uv Timer; Resetting The Cleaning Timer - Thermo Scientific Barnstead Nanopure Life Science 2114 Series Operating Manual And Parts List

Uv/uf ultrapure water system
Table of Contents

Advertisement

Selecting the Set
Point (continued)
Resetting the
Cleaning Timer
Resetting the UV
Thermo Scientific
5. Press ENTER.
6. Press the UP or DOWN arrow until the desired value is displayed.
7. Press ENTER.
8. Press BACK to return to the (Idle) mode.
The second line of the display will show a "Below Set point" message once
every 8 to 10 seconds when the measurement is below the set point. (The
set point indication is set at the factory at 10 MΩ-cm.)
1. From the (Idle) mode, when the display reads, "Nanopure (Idle)
xx:xx:xx xx/xx/xx," press the UP or DOWN arrow until the display
reads, "Reset Timers?"
2. Press ENTER. Display will read, "Reset UV Timer (NO)."
3. Press ENTER.
4. Display will read, "Reset Cleaning Timer" and show on the right side
of the second line when the user will start receiving "Clean Unit"
reminders.
5. Press the UP or DOWN arrow to select "Yes."
6. Press ENTER and the timer will be reset. This will reset the cleaning
timer for approximately six months calendar time.
With the Nanopure in an (Idle) position, the UV timer can be reset by
performing the following steps.
Timer
Note When the user begins receiving "Change UV Lamp" reminders, it
will be time to change UV lamp and reset UV timer as described below.
Section 9
User Settings
Nanopure Life Science (UV/UF)
s
9-5