To Reset Carbon Prefilter Timer; Power Connections - Thermo Scientific EASYpure RoDi Operating Manual

Ultrapure water purification system
Table of Contents

Advertisement

Section 5
Installation

Power Connections

To Reset Carbon
Prefilter Timer
5-8
EASYpure RoDi
1. The power cord connection is located on the upper right corner in the
rear of the unit (see Figure 5-8).
2. Determine which power cord you need (this will be based on your
country and outlets available in your lab). A 120V power cord with (2)
1.6 amp fuses and 240V power cords with (2) 0.63 amp fuses are
provided with the unit.
3. Remove the fuse drawer, install the fuses included with the power cord
to be used, and reinstall drawer.
4. Verify power switch is turned off and attach receptacle end of power
cord into the power socket.
5. Plug other end of power cord into facility power.
Warning Power unit OFF before unplugging unit.
Fuse Drawer
Main
Power Switch
Power Cord
Receptacle
Figure 5-8. Power Connections
The carbon prefilter timer records service time to ensure the reliability of
the RO membrane. It must be reset prior to initial operation and after
cartridge replacement.
1. Turn power OFF using switch on the back of the unit.
2. Press and hold the FLUSH key while turning the unit back ON.