Operating Basics
2 48
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
19. Output the sequence and observe the pattern using an oscilloscope. Refer to
Figure 2-25 on page 2-36 for the connections between the DG2040 Data
Generator and the oscilloscope.
Press START/STOP button on the front-panel to start the output. The
message, Waiting for Trigger, is displayed in the upper right part of the
screen.
The first two lines in the sequence wait for a trigger event. Press the
FORCE TRIGGER button on the front-panel to generate the trigger event.
The sequence in line 2 outputs the BK1 pattern repeatedly until the event
condition is satisfied. Press the STEP/EVENT button on the front-panel to
quit the loop and to advance to sequence line 3.
In Enhanced mode, the entire sequence is repeatedly output. So the message
Waiting for Trigger is displayed again and again until you press the
START/STOP button on the front-panel.
Figure 2 33: Sample sequence
20. Exit sequence creation mode and save the data in a file. Name the file SEQ1.
Menu button
Bottom button
Insert a formatted 3.5 inch floppy disk into the 3.5 inch floppy disk drive.
File
pop up menu
Side button
Save Data &
Setup
DG2040 Data Generator
Front panel