GPIB . The GPIB function sets the GPIB operating mode and address of the
DG2040 Data Generator for remote control or hardcopy.
Base Menu
Item
Function
Configure
Sets the GPIB operating mode.
Operating Mode
Talk/Listen
Talk Only
Off Bus
Address
Sets the instruments GPIB address. The address can be set to a value
between 0 and 30.
Operation.
Do the following to set the GPIB operating mode.
Bottom button
Pop up menu
System
Select GPIB Configure with the up and down
arrow buttons.
Select the operating mode with the left and
right arrow buttons.
Operation.
Do the following to set the GPIB address.
Bottom button
Pop up menu
System
Select GPIB Address with the up and down
arrow buttons.
Select the address with the general purpose
knob.
Reference
Function
Normal remote control
Used for hard copy output.
The connection between the
instrument and the bus is set to
the disconnected state.