Tektronix DG2040 User Manual page 236

Data generator
Hide thumbs Also See for DG2040:
Table of Contents

Advertisement

Test Record
DG2040 Data Generator
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
3. Use the sampling oscilloscope measurement functions to perform the clock
signal rise and fall time measurement under the clock frequency settings
shown in Table B-8.
4. Change the SMA cable connection from the CLOCK and CLOCK outputs to
CH0 and CH0 outputs, respectively, and repeat step 3 for each of these
channels. Verify that the rise and fall times are within the specified ranges.
5. Change the SMA cable connection from the CH0 and CH0 outputs to CH1
and CH1 outputs, respectively, and repeat step 3 for each of these channels.
Verify that the rise and fall times are within the specified ranges.
Table B 8: Rise and fall time accuracies
Voltage settings
Low level
High level
0 V
+1.00 V
0 V
+1.00 V
0 V
+1.00 V
6. Press the START/STOP button on the front panel to stop output (the
button's LED goes off).
This completes the performance verification procedures.
Photocopy the following 4 pages and use them to record the performance test
results for your DG2040 Data Generator.
Appendix B: Performance Verification
Clock frequency
Rise and fall time limit
100 MHz
t150 ps (20 % - 80 %)
500 MHz
t150 ps (20 % - 80 %)
1100 MHz
t150 ps (20 % - 80 %)
B 43