Table B 7: Clock output voltage accuracy
Settings
Low voltage level (V)
High voltage level (V)
-1.125
-0.875
+1.000
+1.25
+3.250
+3.50
-1.125
+1.375
0.000
+2.50
+1.000
+3.50
4. Press the START/STOP button on the front panel to stop output. The LED,
corresponding to the button, is turned off.
Appendix B: Performance Verification
Low level output voltage ranges
0.250 V, ± 0.062 V
(0.188 V to 0.312 V)
0.250 V, ± 0.062 V
(0.188 V to 0.312 V)
0.250 V, ± 0.062 V
(0.188 V to 0.312 V)
2.50 V, ± 0.175 V
(2.325 V to 2.675 V)
2.50 V, ± 0.175 V
(2.325 V to 2.675 V)
2.50 V, ± 0.175 V
(2.325 V to 2.675 V)