Appendix B: Performance Verification
Internal Trigger Generator
& External Trigger Input
B 26
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
This test verifies that the internal trigger generator and the external trigger input
are functioning correctly.
The following equipment is required.
Equipment Required
Connections. Refer to Figure B-14 for connections.
Use the BNC cables to make the following connections:
H Connect the sync output from the DG2040 Data Generator rear panel to the
oscilloscope CH4 input.
H Connect the function generator output to the DG2040 Data Generator front
panel trigger input.
AFG320
CH1
DG2040 Data Generator
Figure B 14: Internal trigger generator & external trigger input connection
Oscilloscope
Function generator
Two BNC cables
Performance check disk
BNC Cable
Trigger Input
BNC Cable: connect to the AUX OUTPUT: SYNC
connector on the DG2040 Data Generator rear panel.