Tektronix DG2040 User Manual page 237

Data generator
Hide thumbs Also See for DG2040:
Table of Contents

Advertisement

Appendix B: Performance Verification
DG2040 Test Record
Serial Number:
Calibration Date:
Table B 9: DG2040 test record
Clock frequency tests
Internal clock frequency
1.100 000 GHz
accuracy
acc ac
650.100 0 MHz
200.000 0 kHz
100.000 0 mHz
High level output voltage accuracy (DC Data
Out)
-0.875 V
0 V
CH0 Output
+1.00 V
+2.00 V
+3.50 V
-0.875 V
0 V
CH0 Output
+1.00 V
+2.00 V
+3.50 V
-0.875 V
0 V
CH1 Output
+1.00 V
+2.00 V
+3.50 V
-0.875 V
0 V
CH1 Output
+1.00 V
+2.00 V
+3.50 V
B 44
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Certificate Number:
Technician:
Low limit
Test result
1.099 998 9 GHz
650.099 35 MHz
199.999 80 kHz
99.999 90 mHz
Low limit
Test result
-0.951 V
-0.050 V
0.920 V
1.890 V
3.345 V
-0.951 V
-0.050 V
0.920 V
1.890 V
3.345 V
-0.951 V
-0.050 V
0.920 V
1.890 V
3.345 V
-0.951 V
-0.050 V
0.920 V
1.890 V
3.345 V
High limit
1.100 001 1 GHz
650.100 65 MHz
200.000 20 kHz
100.000 10 mHz
High limit
-0.799 V
0.050 V
1.080 V
2.110 V
3.655 V
-0.799 V
0.050 V
1.080 V
2.110 V
3.655 V
-0.799 V
0.050 V
1.080 V
2.110 V
3.655 V
-0.799 V
0.050 V
1.080 V
2.110 V
3.655 V
DG2040 Data Generator

Advertisement

Table of Contents
loading

This manual is also suitable for:

Dg203059021-1059021-15

Table of Contents