Setup. Set the oscilloscope and AFG320 Function Generator as indicated below:
H Oscilloscope
Displayed channel
Vertical axis
. . . . .
Horizontal axis
. . .
Record length
. . . .
Acquire mode
. . . .
Acquire sequence
Trigger mode
. . . .
Trigger level
. . . . .
Trigger source
. . . .
Trigger slope
. . . . .
Trigger position
Input coupling
. . . .
Input impedance
InstaVu
. . . . . . . . .
H Function generator
Output channel
. . .
Waveform
. . . . . . .
Frequency
. . . . . . .
Amplitude
. . . . . . .
Offset
. . . . . . . . . .
Procedure. Follow the steps below to verify that the Internal Trigger Generator
and the External Trigger Input are functioning correctly:
1. Load the TP6GEN.PDA test pattern file from the performance check disk.
2. Press the START/STOP button on the front panel. The LED, corresponding
to the button, should be on.
3. Verify that the sync signal appears every 1 ms on the oscilloscope screen.
4. Press the following buttons to set the DG2040 Data Generator trigger
interval to 10 s.
Menu button
Bottom button
SETUP
Trigger
Appendix B: Performance Verification
CH4
5 V/div for CH4
1.00 ms/div
500
Peak Detect
.
RUN/STOP button only
Auto
2 V
CH4
Falling Edge
. .
50 %
DC
. .
1 MW for CH4
off
CH1
Square wave
500 Hz
1 V (50 W termination))
500 mV
Pop up menu
Side button
Interval
State (Set to On)
Time
Front panel
button
10
Hz/s/V
B 27