Tektronix DG2040 User Manual page 205

Data generator
Hide thumbs Also See for DG2040:
Table of Contents

Advertisement

Appendix B: Performance Verification
B 12
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Setup. Set the oscilloscope and AFG320 Function Generator as indicated below:
H Oscilloscope
Displayed channel
Vertical axis
. . . . .
Horizontal axis
. . .
Record length
. . . .
Acquire mode
. . . .
Acquire sequence
Trigger mode
. . . .
Trigger level
. . . . .
Trigger source
. . . .
Trigger position
Trigger slope
. . . . .
Input coupling
. . . .
Input impedance
H Function generator
Output channel
. . .
Waveform
. . . . . . .
Frequency
. . . . . . .
Amplitude
. . . . . . .
Offset
. . . . . . . . . .
Procedure. Follow the steps below to verify pattern data output:
1. Load the TP2PG.PDA test pattern file from the performance check disk.
2. Press the START/STOP button on the front panel. The LED, corresponding
to the button, should be on.
3. Verify that the data pattern shown in Figure B-5 appears on the oscilloscope.
4. Disconnect the SMA cable from the CH0 output connector and the CH0
output connector on the DG2040 Data Generator , and then connect them to
the CH1 output connector and the CH1 output connector on the DG2040
Data Generator.
5. Verify that the pattern data shown in Figure B-5 appears on the oscilloscope.
6. Disconnect the BNC cable from event input connector on the DG2040 Data
Generator rear panel.
7. Change the oscilloscope acquisition sequence to Single Acquisition
Sequence.
CH1, CH2, CH3 and CH4
1 V/div for CH1 and CH2
5 V/div for CH3 and CH4
5 ms/div
500
Peak Detect
.
RUN/STOP button only
Auto
2 V
CH3
. .
50 %
Falling Edge
DC
50 W for CH1 and CH2
. .
1 MW for CH3 and CH4
CH1
Square wave
500 Hz
1 V (50 W termination))
500 mV
DG2040 Data Generator