Appendix B: Performance Verification
Maximum Operating
Frequency (Trigger
Operation)
B 22
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
This test verifies that the maximum operating frequency (trigger operation) is
functioning correctly.
The following equipment is required.
Equipment Required
Connections. Refer to Figure B-11 for connections.
Use the SMA cables and the SMA female-to-BNC male adapters to make the
following connections:
H Connect CH0 output from the DG2040 Data Generator front panel to the
oscilloscope CH1 input.
H Connect CH1 output from the DG2040 Data Generator front panel to the
oscilloscope CH2 input.
H Use the remaining two SMA cables, the remaining two SMA female-to-BNC
male adapters, and the 50 W terminations to terminate the CH0 and CH1
outputs.
Use the BNC cables to make the following connections:
H Connect the function generator CH1 output to the DG2040 Data Generator
rear panel event input.
H Connect the event output from the DG2040 Data Generator rear panel to the
oscilloscope CH3 input.
H Connect the sync output from the DG2040 Data Generator rear panel to the
oscilloscope CH4 input.
Oscilloscope
Function generator
Three BNC cables
Four SMA cables
Four SMA female to BNC male adapters
Performance check disk
DG2040 Data Generator