This test verifies maximum operating frequency (continuous) is functioning
correctly.
The following equipment is required.
Equipment Required
Connections. Refer to Figure B-7 for connections.
Use the SMA cables and the SMA female-to-BNC male adapters to make the
following connections:
H Connect CH0 output from the DG2040 Data Generator front panel to the
oscilloscope CH1 input.
H Connect CH1 output to the DG2040 Data Generator front panel to the
oscilloscope CH2 input.
H Use the remaining two SMA cables, the remaining two SMA female-to-BNC
male adapters, and the 50 W terminations to terminate the CH0 and CH1
outputs.
Use the BNC cables to make the following connections:
H Connect the function generator Ch1 output to the DG2040 Data Generator
rear panel event input.
H Connect the event output from the DG2040 Data Generator rear panel to the
oscilloscope CH3 input.
H Connect the sync output from the DG2040 Data Generator rear panel to the
oscilloscope CH4 input.
Appendix B: Performance Verification
Oscilloscope
Function generator
Three BNC cables
Four SMA cables
Four SMA female to BNC male adapters
Performance check disk
B 17